1010 Execution Error: Vacuum pressure (IF) is too high.
The gate valve cannot be opened because the vacuum pressure (IF/BK) exceeds the limit (When the instrument is changing from Standby mode to Analysis Mode).
Check the vacuum pressure (IF/BK) for the following possible issues:
- The orifice of the sampling cone is too big.
- The seal of the sampling cone is incomplete (Defective gasket).
- The plasma is too far from the sampling cone.
- The power of the foreline pump is not connected or the breaker is OFF.
- The foreline pump oil level is low or the oil is dirty.
- The vacuum tubing has failed (the tubing is off, or a leak is caused by an incomplete seal such as a pin hole or a defective O-ring).
- A drop of water was introduced into the vacuum chamber.
- There is a fault in the Pirani gauge (IF/BK) or the Connection PCA.