1063
Error: ADS 2 Switching valve (Valve A or B ) error.
The ADS 2 switching valves (Valve A or B) has reported and error.
The following causes are inferred:
- The instrument operation has become unstable, Power Cycle and Initialize
the ADS 2.
- The valve (A or B) is not installed.
- The valve (A or B) is not secured correctly.
- The valve (A or B) drive coupling is not installed.
- The valve (A or B) has seized or is obstructed in it's rotation.
Check and clean the valve rotor and stator
- The incorrect valve type has been fitted.
- The valve drive motor has failed.
Once corrected Power Cycle and Initialize the ADS 2 and Initialize.