1065
Error: ADS 2 Communication error.
Communication with the ADS 2 has failed.
The following causes are inferred:
- The ADS interface cable is not connected. Check and confirm
the ADS 2 connection cable is correctly fitted.
- The instrument operation has become unstable, Power Cycle and Initialize
the ADS 2 .
- There is a fault with the ADS 2 module control PCA (board).