1067
Error: ADS 2 Firmware mismatch error.
The ADS 2 firmware does not match the expected version required by ICPMS
MassHunter.
The following causes are inferred:
- The instrument operation has become unstable, Power Cycle and Initialize
the ADS 2 .
- Verify the version of MassHunter is compatible with the AVS 2.