1120 Error: Vacuum pressure (IF) is too high to stay in Analysis mode.
Analysis Mode cannot be maintained because the vacuum pressure (IF/BK) exceeds the
limit (In Analysis Mode).
Check the vacuum pressure (IF/BK) for the following possible issues.
- The diameter of the hole in the sampling cone is too big.
- The seal of the sampling cone is incomplete (Defective gasket).
- The plasma is too far from the sampling cone.
- The power of the foreline pump is not connected or the breaker is OFF.
- The foreline pump oil level is low or the oil is dirty.
- The vacuum tubing has failed (the tubing is off, or a leak is caused by an incomplete seal such as a pin hole or a defective O-ring).
- A drop of water was introduced into the vacuum chamber.