1123   Error: Vacuum pressure (AN) time out. Vacuum too high to enter Standby mode.

The instrument cannot transit to Standby mode because the vacuum pressure (AN) is not below the limit within the prescribed time (When the instrument is changing from Shutdown mode to Standby mode).

Check the pressure of the analyzer chamber for the following possible issues. Also this error is generated when a vacuum pump is not used for a long time. In this case, turn on the vacuum ON again.