1166   Error: HV for Ion gauge is out of range (Standby).

The Ion gauge driver for the Connection PCA reads that 180V High-voltage is not

supplied from the HV driver module, and the Main PCA determines that 180V

High-voltage is not being supplied (In Standby mode).

This error is displayed for more than 210V or less than 150V.

The following are the probable causes.