1309 Error: OctPole control malfunction.
The RF voltage of the Octopole cannot be controlled.
Check the RF voltage of the Octopole. The following are the probable causes.
- 48V is not supplied (Cover is open).
- Octopole rod or reaction cell-related wiring (including the vacuum chamber) causes a short in other metallic parts.
- Octopole matching frequency is not appropriate.
- Fault of the ORS cell driver PCA.