1440 Error: Cooling water (IF) flow rate is too low. Spray chamber temperature control has been turned off.
The flow rate of the cooling water (RF/WC/IF) is below the limit, and the cooling of the spray chamber is stopped.
Check the flow rate of the cooling water. The following are the probable causes.
- The power of the cooling water supply is not switched ON.
- The cooling water valve is closed.
- The cooling water is insufficient (possibility of a water leak).
- Clogged cooling water tubing and a leak.
- The cooling strainer is clogged.
- Faulty cooling water supply.
- Faulty Introduction PCA.