2150 Warning: Interface Vacuum is too low.
The vacuum pressure (IF/BK) is below the limit (In Analysis Mode).
Check the vacuum pressure (IF/BK) for the following possible issues:
- The orifice of the sampling cone is too small or clogged.
- The orifice of the skimmer cone is too small or clogged.