2152 Warning: Analyzer Vacuum is too high.
The vacuum pressure (AN) exceeds the limit (In Analysis Mode).
Check the vacuum pressure (AN) for the following possible issues:
- The orifice of the sampling cone is too big.
- The seal of the sampling cone is incomplete (Defective gasket).
- The plasma is too far from the sampling cone.
- The power of the foreline pump is not connected or the breaker is OFF.
- The foreline pump oil level is low or the oil is dirty.
- The vacuum tubing has failed (the tubing is off, or a leak is caused by an incomplete seal such as a pin hole or a defective O-ring).
- A drop of water was introduced into the vacuum chamber.
- The vent valve is not closed or the seal is incomplete.
- The cover of the vacuum chamber is not closed or the seal is incomplete.
- There is a leak in the gate valve.
- There is a fault in the Ion gauge or the Connection PCA.
- The cell gas flow rate is too high.