The data analysis method table displays various tables for each method development task. This section describes the following operations in the Data Analysis method table:
You can fill a column of cells with values and units, or change the entire column. To fill values, follow these steps. Use the same procedure to change the column in a single step.
The entire column is selected and the first cell changes to green.
The first cell is filled with a number or unit.
All cells in the column will be filled with the value in the first cell.
When you change all the units in the column, the change is not made if the unit in the first cell is the same. In this case, select a different unit and change the column to that unit first. Then change the column to the desired unit.
Use this procedure to fill or change all concentrations for a calibration curve level in a column. To automatically set the concentration, enter a multiplier for each calibration curve level.
In this example, we change the concentrations for levels 1 to 4 from “0”, “10”, “20”, and “50” to “0”, “10”, “25”, and “100”.
The entire column is selected.
Concentration levels that are set to “0” cannot be used as a reference.
To enter a new concentration or change the concentration of a reference level, first fill the column with values as described in “Filling values and units”.
The [Multiply Conc] dialog box appears.
[Multiply Conc] Dialog Box
A “1” is displayed in the level to be used as the reference. The other levels contain the current multiplier.
In our example, the following values would be entered.
[Multiply Conc] Dialog Box
The concentration for each calibration curve level changes.
You can sort the method table by Tune Mode or Mass by using the headers. To sort the table, follow these steps:
The method table will be sorted in ascending or descending order.
:Descending order.
:Ascending order.
:Descending order.
:Ascending order.
To display the table in the original sort order, select <Reset Sort> from the context menu.