Item |
Description |
|---|---|
Summary |
7700x Generic Method for high matrix samples |
Compatible Sample Types |
Generic Method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7700x, due to the use of HMI aerosol dilution and the short pre-defined cell gas stabilization time. |
Required Hardware |
7700x only, HMI, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7700x Generic Method for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7700x, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7700x only, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7700x Generic Method for medium matrix samples |
Compatible Sample Types |
Generic Method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7700x, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7700x only, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7700s Generic Method with high sensitivity for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous samples (e.g. 1%nitric acid solution), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7700s, due to the use of H2 cell mode and cool plasma conditions for some analytes. |
Required Hardware |
7700s, s-Lens, Micro Flow Nebulizer |
Item |
Description |
|---|---|
Summary |
8800 Generic Method for high matrix samples |
Compatible Sample Types |
Generic Method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 8800 ICP-MS Triple Quad. |
Required Hardware |
8800 ICP-MS Triple Quad, HMI, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8800 Generic Method for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 8800 ICP-MS Triple Quad. |
Required Hardware |
8800 ICP-MS Triple Quad, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8800 Generic Method for medium matrix samples |
Compatible Sample Types |
Generic Method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 8800 ICP-MS Triple Quad. |
Required Hardware |
8800 ICP-MS Triple Quad, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8800 Generic Method with high sensitivity for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous samples (e.g. 1%nitric acid solution), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 8800 ICP-MS Triple Quad, due to the use of H2 cell mode and cool plasma conditions for some analytes. |
Required Hardware |
8800 ICP-MS Triple Quad, s-Lens, Micro Flow Nebulizer |
Item |
Description |
|---|---|
Summary |
7900 Generic Method for high matrix samples |
Compatible Sample Types |
Generic Method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7900, due to the use of HMI aerosol dilution and the short pre-defined cell gas stabilization time. |
Required Hardware |
7900, HMI, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7900 Generic Method for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7900, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7900, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7900 Generic Method for medium matrix samples |
Compatible Sample Types |
Generic Method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7900, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7900, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7900 Generic Method with high sensitivity for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous samples (e.g. 1%nitric acid solution), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7900, due to the use of H2 cell mode and cool plasma conditions for some analytes. |
Required Hardware |
7900, s-Lens, Micro Flow Nebulizer |
Item |
Description |
|---|---|
Summary |
7800 Generic Method for high matrix samples |
Compatible Sample Types |
Generic Method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7800, due to the use of HMI aerosol dilution and the short pre-defined cell gas stabilization time. |
Required Hardware |
7800, HMI, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7800 Generic Method for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7800, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7800, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7800 Generic Method for medium matrix samples |
Compatible Sample Types |
Generic Method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7800, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7800, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
8900 Generic Method for high matrix samples |
Compatible Sample Types |
Generic method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Tune : UHMI-8 Cell mode : No gas, H2, He, HEHe, O2, NH3 |
Required Hardware |
8900 ICP-MS Triple Quad, HMI, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8900 Generic Method for low matrix samples |
Compatible Sample Types |
Generic method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Tune : Low Matrix Cell mode : No gas, H2, He, HEHe, O2, NH3 |
Required Hardware |
8900 ICP-MS Triple Quad, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8900 Generic Method for medium matrix samples |
Compatible Sample Types |
Generic method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where high matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Tune : General Purpose Cell mode : No gas, H2, He, HEHe, O2, NH3 |
Required Hardware |
8900 ICP-MS Triple Quad, x-Lens, MicroMist Nebulizer |
Item |
Description |
|---|---|
Summary |
8900s Generic method for low matrix aqueous sample with H2 and/or NH3 |
Compatible Sample Types |
Generic method for low matrix aqueous samples (e.g. 1%nitric acid solution), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 8900 ICP-MS Triple Quad, due to the use of H2 and/or NH3 cell mode and cool plasma conditions for some analytes. |
Required Hardware |
8900 ICP-MS Triple Quad, s-Lens, 2.5mm injector torch, quartz spray chamber, quartz connector tube(straight), MicroFlow Nebulizer, Pt sampling cone for s-lens, Pt skimmer cone for s-lens, Brass skimmer base for s-lens. |
Item |
Description |
|---|---|
Summary |
m-Lens application (K and Na BEC <10 ppt) |
Compatible Sample Types |
m-Lens application. TDS < 500 ppm is recommended. |
Pre-Defined Analytes |
Li, Be, B, Na, Mg, Al, Si, P, S, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Ni, Co, Cu, Zn, Ga, Ge, As, Se, Br, Rb, Sr, Y, Zr, Nb, Mo, Ru, Rh, Pd, Ag, Cd, In, Sn, Sb, Te, I, Cs, Ba, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, Hf, W, Ta, Re, Os, Ir, Pt, Au, Hg, Tl, Pb, Bi, Th, U |
Comment |
m-Lens application ( hot only mode ). TDS < 500 ppm is recommended. BEC for K and Na < 10ppt. More stable than "high-sensitivity" method with about 4 times lower sensitivity. If concomitant Si is high P must be measured in HEHe or H2 mode as PH4+. |
Required Hardware |
8900 ICP-MS Triple Quad, m-Lens, 2.5mm injector torch, Micro Flow Nebulizer with self aspiration |
Item |
Description |
|---|---|
Summary |
m-Lens application (K and Na BEC <20 ppt) |
Compatible Sample Types |
m-Lens application. TDS < 500 ppm is recommended. |
Pre-Defined Analytes |
Li, Be, B, Na, Mg, Al, Si, P, S, K, Ca, Sc, Ti, V, Cr, Mn, Fe, Ni, Co, Cu, Zn, Ga, Ge, As, Se, Br, Rb, Sr, Y, Zr, Nb, Mo, Ru, Rh, Pd, Ag, Cd, In, Sn, Sb, Te, I, Cs, Ba, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, Lu, Hf, W, Ta, Re, Os, Ir, Pt, Au, Hg, Tl, Pb, Bi, Th, U |
Comment |
m-Lens application ( hot only mode ). TDS < 500 ppm is recommended. BEC for K and Na < 10-20ppt. If concomitant Si is high P must be measured in HEHe or H2 mode as PH4+. |
Required Hardware |
8900 ICP-MS Triple Quad, m-Lens, 2.5mm injector torch, Micro Flow Nebulizer with self aspiration |
Item |
Description |
|---|---|
Summary |
7850 Generic Method for high matrix samples |
Compatible Sample Types |
Generic Method for high matrix aqueous or acid digested samples, and where exceptionally high matrix tolerance is essential. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7850, due to the use of HMI aerosol dilution and the short pre-defined cell gas stabilization time. |
Required Hardware |
7850, HMI, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7850 Generic Method for low matrix samples |
Compatible Sample Types |
Generic Method for low matrix aqueous or acid digested samples (<0.1% Total Dissolved Solids), and where high sensitivity is required. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7850, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7850, x-Lens, Micro Mist Nebulizer |
Item |
Description |
|---|---|
Summary |
7850 Generic Method for medium matrix samples |
Compatible Sample Types |
Generic Method for typical aqueous or acid digested samples (up to 0.2% Total Dissolved Solids), and where matrix tolerance is more important than high sensitivity. |
Pre-Defined Analytes |
None specified |
Comment |
Method is suitable only for the 7850, due to the short pre-defined cell gas stabilization time. |
Required Hardware |
7850, x-Lens, Micro Mist Nebulizer |