Internal Standard Method and Standard Addition Method

Internal Standard Solution

This method is recommended in the following cases:

In the calibration report or the calibration graphs, the ratio of the analyte to internal standard is given, not the counts.

Matrix effects are not normally an issue with drinking water, but the use of internal standards is recommended if the water hardness is high. The method of standard additions can also be used for this purpose.

Standard Addition to a Sample

This method is recommended to correct matrix effects such as signal suppression or enhancement. A standard solution is directly added to a sample. Since a small amount of the standard solution is added to the sample, the matrix effect is the same for both the sample and the standard. This effect can be used to correct for signal suppression or enhancement.

When the signals of sample and spiked solutions are s0, s1 and s2, and the signal of a blank is bkg, the calibration curve obtained by a least squares fit of the points is shown below. Using the bkg may be a problem if the sample matrix is quite different from the bkg matrix.

The concentration of analyte added is p1 and p2, and consequently c0 is the concentration of analyte in the sample. When the concentration of the sample is c0, the actual analyte concentrations in the spiked solutions are c1 and c2; e.g. when 10 ppb (p1) and 20 ppb(p2) are added to the sample and the sample concentration is 2 ppb(c0), the actual concentrations of the spiked samples are 12 ppb (c1) and 22 ppb (c2).

Illust_00253_A_200616.jpg 

Method of Standard Additions

Once a standard addition calibration curve is made for one sample, this calibration curve can be converted to an external calibration curve. After conversion, the vertical axis shifts to the bkg point and the point at the concentration of the background becomes the y-intercept. The concentration of the background becomes 0 and the concentrations of other points are recalculated relative to the background; concentrations of other points are combined values of spiked sample concentrations and sample concentrations. After the conversion, the calibration point of a 10 ppb spiked solution becomes 12 ppb when using the above example. The calibration table is also updated.

If the matrix of other samples is similar to the first sample, the converted external calibration curve can be used.