New features in MassHunter 4.2 make creating and calibrating a chromatographic data analysis method easier than ever.
As the first step, you need to acquire a set of calibration standards and ensure that the resolution, sensitivity, peak shape, baseline etc. are acceptable. In general, this can be verified visually.
If you are working with a previously acquired batch of samples, you can open the batch using Offline Data Analysis by clicking [Open Folder] from the [Batch] group on the [Home] tab, and selecting the batch from the list. The list of samples in the batch is displayed in the Batch at a Glance table.
Ensure that the calibration standards are designated as sample type “Calstd” in Data Analysis. If they were originally acquired as sample type “Sample” during acquisition, change the sample type in the Batch View Grid. You do not need to include a Calblk in chromatographic analysis. The lowest concentration standard should contain recognizable peaks for all analytes.
The [Method Editor] window is displayed.
The total ion chromatogram (TIC) for the selected calibration standard data file is displayed as shown below. If your acquisition includes more than one element or isotope and internal standards, the baseline may look noisy and the peaks may appear smaller than expected because you are seeing a composite of all signals.
For example, if you are looking for arsenic compounds, select mass 75.
The integrated extracted ion chromatogram (EIC) for the selected mass is displayed.
If your method includes compounds or species of different elements, then select <All> from the TIC/EIC menu to display all acquired EICs and the TIC simultaneously. This allows you to add all your selected analyte peaks from different masses to the peak list table graphically. You may need to enlarge the Chromatogram window to see all of the EIC traces clearly if there are more than 2-3. In the example below, the only analyte is As at m/z = 75. However, m/z 35 was included to indicate the presence of interferences from ArCl due to Cl containing peaks.
The manually integrated peaks are red.
Before manual integration above; after manual integration below.
The integrated peaks are added to the Peak list in order of retention time.
Alternatively, you can choose individual peaks from each EIC to
add to the Peak Table by using the “Add Peak Mode” tool .
At this point, you can edit the peak table to change the compound names, designation as analyte, or internal standard and target or qualifier ion. If the auto-integration process did not correctly, integrate a peak (in this example peak 5 was incorrectly integrated due to a noisy apex), select the Smoothing column for the noisy peaks, and then turn on Gaussian smoothing.
The windows should be wide enough to easily capture the peak start and end, but not so wide as to contain several peaks (if possible). Adjust the integration windows as needed by changing the “Integ Start” and “Integ End” values in the Peak Table. The graphic below on the left shows a correctly set integration window for peak 1. The graphic on the right shows an integration window for peak 2 that is too wide.
Results are displayed for the entire batch via the Batch at a Glance
Table at the top, and graphically for each sample at the bottom. Click
individual samples and/or peaks from the table to select them. The
Chromatogram pane shows the EIC for the entire run on top, and the
selected analyte peak on the bottom. If the baseline on the target
ion (bottom chromatogram trace) indicates incorrect integration, use
the manual integration tool to correct
the integration. The [Process Batch] icon changes to
.
Click this icon to requant the batch including your manually integrated
peaks.