Workflow Examples

The following examples go through some of the basic parts of the software to give some basic software training.

Soils (IAEA SOIL-7)

  1. Set the acquisition parameters as follows:

    Measurement m/z number --- all

    Peak pattern --- Semi Quant (6 points/mass) to check for spectral interferences

    Integration time --- 0.05 sec

    Repetition --- 1

  2. Analyze a blank (1% HNO3) first, then analyze a 10 ppb standard solution (normally used for tuning).
  3. Correct for the response factor for semiquantitative analysis using the file for the standard solution, and use the file for the blank solution as the background file.
  4. Analyze the sample, then identify some peaks using the [Template] and label them, then get the semiquantitative result in [Data Analysis].

    The following items can also be selected in SemiQuant.

    • Internal standard correction
    • Subtraction of the blank data file

 

Only the highest point of the 6 points is used for semiquantitative analysis. Therefore the counts in Tabulate/mass and those in the semiquantitative report will be different.

 

  1. Based on the semiquantitative result, quantitate by the method of standard additions as follows:

    Solutions --- pure water, sample, sample + STD 1, and sample + STD 2 (Select the appropriate concentrations for STD 1 and 2 based on the semiquant results)

    Analytes --- V, Cr, Co, Ni, Cu, Zn, As, Cd, Sb, and Pb

  2. Set the acquisition parameters as follows:

    Measurement mass number and integration time ---Select as appropriate

    Peak Pattern --- Full Quant (3 points/mass)

    Repetition --- 3

  3. Analyze pure water to use for the background data file, then the sample and sample + spike solutions.
  4. Make a calibration curve using Standard Addition.

    Refer to Calibration pane and Calibration Parameters Table.

Trace Elements in Water (NIST SRM 1643)

  1. Make the standard solutions for the calibration curve as follows:

    Standard solutions --- blank, 10, 50 and 100 ppb mixed solution

    Analytes --- Al, V, Cr, Mn, Co, Ni, Cu, Zn, As, Cd, and Pb

    Internal standard elements --- Sc (for Al, V, Cr, Mn, Co, Ni, Cu, Zn and As), In (for Cd) and Ho (for Pb)

  2. Set up the method and the sequence.

    Set the acquisition parameters as follows:

    Measurement mass number and integration time ---Select as appropriate

    Peak Pattern --- Full Quant (3 points/mass)

    Repetition --- 3

  3.  

    Using both interference equations and internal standard correction, at first raw counts are corrected by the interference equation, and then the internal standard correction is performed. Therefore, ISTD selection for masses used in the interference equation (the masses in parentheses) is not necessary in the calibration table.

    Remember, background subtraction subtracts the raw data, before ISTD correction, so it cannot be used for quantitation, since it would subtract the ISTD counts from the sample. Always use blank subtraction for subtracting a reagent blank during quantitation. This method only makes the subtraction after the ISTD correction has been applied to both blank and sample.

     

  4. Analyze the standard solutions and sample, and then generate the quantitative results.

Analysis of Semiconductor Samples

Since ICP-MSICP-QQQ allows ppt or sub-ppt level elemental analysis, it has been commonly used in the semiconductor industry to determine ultra trace levels of impurities in chemicals and semiconductor devices. In this section, the analysis of semiconductor samples is described.

General Considerations

Environment

In semiconductor devices, contamination deteriorates performance: alkali and alkali-earth elements reduce breakdown voltage, transition metals reduce the lifetime of the carrier causing higher dark current, doping elements cause a shift in the device operating point and particles cause short circuits. Therefore, this contamination has to be controlled in the semiconductor industry. However, these contaminants exist everywhere in the atmosphere and special attention has to be paid to prevent airborne contamination of the various chemical and semiconductor device samples to be analyzed by ICP-MSICP-QQQ. The best way to avoid such contamination is to install the ICP-MSICP-QQQ in a clean room.

This instrument is designed for clean room use. The foreline pumps can be installed outside the clean room while the instrument itself is installed inside the clean room. All cooling air flows from outside to inside and is vented through the ventilation duct at the top of this instrument, which prevents introduction of particles from the instrument into the clean room.

The cleanliness of a clean room is expressed by its class number. Class 1000 means that the number of particles bigger than 0.5 mm is 1000 per cubic foot. The normal environment is worse than class 1,000,000. To get better cleanliness, the floor of the clean room must have a grating to allow air to flow down. Installing this instrument in a class 1000 clean room is recommended.

For the sample and standard solution preparation facility, better cleanliness will be required: a class 100 clean room with a clean draft.

Clinical Sample Preparation

This is intended as a quick start guide for liquid clinical sample preparation and analysis by this instrument. It includes "recipes" for sample dissolution and general considerations for trace clinical determination. It is not intended as an exhaustive source of information or critical sample preparation review.

Reagents & Reagent Selection

Because reagent selection is at least as important as correct sampling procedures it is imperative that all reagents are of an appropriate high quality. The majority of clinical matrices can be prepared in a basic solution and we will focus mainly on these reagents. The following text lists suggested reagents and their CAS number (please check with your local laboratory reagents supplier).

Water

Reason for use:

Main solvent & diluent base high purity essential

The most important reagent. Use Milli-Q or Elga etc Ultrapure water system (>18 MW)

Ammonium Hydroxide Solution (NH4OH, Ammonia Solution etc.) [CAS 1336-21-6]

Reason for use:

Basic reagent solubilises cells and stops protein precipitation

High purity or Ultra purity solution (preferred).

Butan-1-ol (1-butanol, butanol) [CAS 71-36-3]

Reason for use:

Acts as a carbon buffer for standards and different clinical matrices improves sensitivity for high I.P. elements (As, Se)

High purity or Ultra purity solution.

EDTA (ethylenediaminetetraacetic acid, H4-EDTA) [CAS 60-00-4]

Reason for use:

Complexing agent stabilizes metals in basic solution

High purity preferred. Order as the ACID (Not Na salt!)

TRITON X-100 (non-ionic surfactant) [CAS 9002-93-1]

Reason for use:

Improves sample introduction wettability and reduces likelihood of precipitation or blockage in spray chamber and torch injector

High purity preferred.

Nitric Acid [CAS 7697-37-2]

Reason for use:

Limited for clinical matrices only useful if urine samples form precipitate due to high phosphate levels. When using acidic media, all calibration and rinse solutions must be acidic in order to keep the chemistry of the sample introduction system stable. An acid concentration of 1% should be sufficient.

Solution Preparation

To prepare the diluent solution, the reagents need to be mixed in a pre-cleaned low-density polyethylene or polypropylene bottle (PFA or FEP can also be used). Mix the components as listed in the table below (with example weights assuming a final solution volume of 1 liter).

Reagent

% in solution (w/v)

Quantity (for 1l solution)

1-butanol

2%

0.71oz

EDTA

0.05%

0.02oz

Triton X-100

0.05%

0.02oz

NH4OH

1%

0.35oz

Water

Make up to volume

Mix the solution thoroughly before use.

Sample Preparation

Samples can be prepared by simple dilution - for blood, serum & plasma, a 10-fold dilution of the sample should be sufficient. For urine, use a minimum 20-fold dilution to keep the salt levels low enough to allow extended batch runs.

Sample preparation example

Sample preparation example when using the Agilent I-AS autosampler configured with the 89-position rack (6ml vials).

Pipette 0.5ml of sample (plasma, serum, blood) or 0.25ml of urine into the I-AS vial using single-use plastic pipette tips then pipette 4.5ml (plasma, serum, blood) or 4.75ml (urine) of the dilution matrix using a 5ml dispensing pipette. Pipetting the larger volume on top of the sample should give sufficient mixing, but further mixing can be done with ultrasonification, if necessary.

The volumes can be adjusted as required due to sample volume limitations. If the system is fitted with a low flow nebulizer, a full element suite can be determined with limited sample volume.

Calibrate the system with standards that are prepared in the diluent solution. All rinses should also use the diluent solution to keep the chemistry of the sample introduction system stable.