Configure the peripump settings.
Select a sample introduction mode to use from the list.
Set the time (sec) in the [Time] column, the speed (rps) in the [Speed Nebulizer Pump] column, the vial position in the [Vial#] column. Gray cells cannot be set.
You can set [HOME] position from the dropdown menu. In other cases, click [<Vial#>] to open a dialog box, and then select a vial in the dialog box. You can directly enter a vial number from the keyboard into the [<Vial#>] column.
The items displayed depend on the sample introduction mode you use.
Skip this step if you enter 0 in Time and Speed or if it is left blank.
Pre Run
[Sample Uptake]
Set the uptake time and uptake speed before acquisition.
[Stabilize]
Set the stabilization time after sample uptake before starting the signal acquisition.
Acquisition
[Speed]
Set the pump speed to use for acquisition.
Post Run
If the Autosampler is SPS 4, Dual Port Rinse is supported. You can select [Rinse Port1] or [Rinse Port2] from the [Vial#] column. When you use Dual Port Rinse, you need to turn on Dual Port Rinse on the Rack tab in the [Configure Autosampler] dialog box.
[Uptake]
Displayed when [Discrete Sampling] (AVS) or [Autodilution] (ADS 2) is selected in [Sample introduction] .
During Post Run, the valve switches to [ISTD].
Set the ISTD solution uptake time and the pump speed.
[Stabilize]
Displayed when [Discrete Sampling] (AVS) or [Autodilution] (ADS 2) is selected in [Sample introduction] .
After uptake, set the stabilization time and the pump speed before starting the Pre Run for the next acquisition.
The following items are not be displayed.
[Probe Rinse]
After the acquisition of the standard solution, enter the rinse time of the probe at the rinse port and the pump speed.
[Rinse1]
Enter the rinse time, pump speed, and vial position for the external bottle 1.
[Probe Rinse1]
After Rinse1, enter the rinse time for the probe at the rinse port and the pump speed.
[Rinse2]
Enter the rinse time, pump speed, and vial position for the external bottle 2.
[Probe Rinse2]
After Rinse2, enter the rinse time for the probe at the rinse port and the pump speed.
[Rinse3]
Enter the rinse time, pump speed, and vial position for the external bottle 3.
[Probe Rinse3]
After Rinse3, enter the rinse time for the probe at the rinse port and the pump speed.
[Rinse4]
This item is displayed when General] is selected in the Sample Introduction.
Enter the rise time, pump speed, and vial position for the external bottle 4.
[Probe Rinse4]
This item is displayed when General] is selected in the Sample Introduction.
After Rinse4, enter the probe rinse time at the rinse port and the pump speed.