[PeriPump Setting] dialog box

Configure the peripump settings.

Sample Introduction

Select a sample introduction mode to use from the list.

Setting Table

Set the time (sec) in the [Time] column, the speed (rps) in the [Speed Nebulizer Pump] column, the vial position in the [Vial#] column. Gray cells cannot be set.

You can set [HOME] position from the dropdown menu. In other cases, click [<Vial#>] to open a dialog box, and then select a vial in the dialog box. You can directly enter a vial number from the keyboard into the [<Vial#>] column.

 

 

Pre Run

[Sample Uptake]

Set the uptake time and uptake speed before acquisition.

[Stabilize]

Set the stabilization time after sample uptake before starting the signal acquisition.

Acquisition

[Speed]

Set the pump speed to use for acquisition.

Post Run

 

If the Autosampler is SPS 4, Dual Port Rinse is supported. You can select [Rinse Port1] or [Rinse Port2] from the [Vial#] column. When you use Dual Port Rinse, you need to turn on Dual Port Rinse on the Rack tab in the [Configure Autosampler] dialog box.

 

[Uptake]

Displayed when [Discrete Sampling] (AVS) or [Autodilution] (ADS 2) is selected in [Sample introduction] .

During Post Run, the valve switches to [ISTD].

Set the ISTD solution uptake time and the pump speed.

[Stabilize]

Displayed when [Discrete Sampling] (AVS) or [Autodilution] (ADS 2) is selected in [Sample introduction] .

After uptake, set the stabilization time and the pump speed before starting the Pre Run for the next acquisition.

The following items are not be displayed.

[Probe Rinse]

After the acquisition of the standard solution, enter the rinse time of the probe at the rinse port and the pump speed.

[Rinse1]

Enter the rinse time, pump speed, and vial position for the external bottle 1.

[Probe Rinse1]

After Rinse1, enter the rinse time for the probe at the rinse port and the pump speed.

[Rinse2]

Enter the rinse time, pump speed, and vial position for the external bottle 2.

[Probe Rinse2]

After Rinse2, enter the rinse time for the probe at the rinse port and the pump speed.

[Rinse3]

Enter the rinse time, pump speed, and vial position for the external bottle 3.

[Probe Rinse3]

After Rinse3, enter the rinse time for the probe at the rinse port and the pump speed.

[Rinse4]

This item is displayed when General] is selected in the Sample Introduction.

Enter the rise time, pump speed, and vial position for the external bottle 4.

[Probe Rinse4]

This item is displayed when General] is selected in the Sample Introduction.

After Rinse4, enter the probe rinse time at the rinse port and the pump speed.

To Get Here

See Also