This dialog box is displayed when [LA] or [other] is selected from the [Sample Introduction] list.
Start Acquisition with Remote Signal
You can select “Use Remote Signal” and then set Delay time for startup. When you do this and add to the Queue you will see the message in the Status bar that the ICP-MS is waiting for the LA instrument until it send the APG remote signal.
Then when the signal is sent, the software will wait the time specified, for stabilization of the LA signal, and the time remaining will appear in the Status bar.
Wait for Ready in 2nd or Later Replicate
If you select “Wait for Ready in 2nd or Later Replicate” data acquisition for the 2nd and later iterations of a repeated analysis starts without waiting for the READY signal. An example application of this option is a repeated analysis using laser ablation.
Remote Shutdown Options
This item is displayed when [other] is selected from the [Sample Introduction] list. Select a delivery action for the Remote Shutdown signal.
Send shutdown signal when plasma is off
Select it when you use a unsupported LC for connection.
Send shutdown signal during plasma is off process
Select it when you use FFF for connection.
<OK>
Applies the settings.