[Set Prohibited Masses] dialog box

Set prohibited masses.

Insert prohibited masses to protect the detector (EM) from large signals caused by Oxygen, Argon, Nitrogen, Hydrogen, and other polyatomic ions.

When it is installed, the following default values are configured.

Lenses

Mode

Prohibited Masses

x-Lens/ s-Lens common

No Gas - Hot Mode

16 17 18 19 40 41

No Gas - Cool Mode

16 17 18 19

H2 Gas Mode

16 17 18 19

He Gas Mode

16 17 18 19

Lenses

Mode

Prohibited Masses

x-Lens/ s-Lens/m-Lens common

No Gas - Hot Mode

16 17 18 19 40 41

No Gas - Cool Mode

16 17 18 19

H2 Gas Mode

16 17 18 19

He Gas Mode

16 17 18 19

3rd Cell Gas

16 17 18 19

4th Cell Gas

16 17 18 19

Setup Table

[Mode] column

Displays each tune mode.

[Masses] column

Displays the specified prohibited masses. They can be changed (see CAUTION note below).

[OK]

Applies the settings.

[Cancel]

Cancels the settings.

 

Do not change the prohibited masses setting except for specific applications, such as measurement of 40Ca with cool plasma conditions).

 

If under normal plasma conditions, large signals will cause the rapid deterioration of the detector (EM).