Set the actions to do while the probe of the autosampler is moving, as necessary.
When prepFAST is used for the Autosampler, the label changes to "prepFAST Dilution".
Stop PeriPump Rotation
The peristaltic pump will stop when the Autosampler probe is in motion, to prevent air from being taken in from the sample probe.
Change Nebulizer Gas Flow Rate
While the Autosampler probe is in motion, the nebulizer gas (also called carrier gas) flow rate changes to the set value. If a stabilization time is set, the probe moves to the specified cleaning port after the set time. When the self-aspiration is used instead of a peristaltic pump, you can use this action to avoid air aspiration while the probe is moving.
[Flow Rate]
Set the flow rate of the nebulizer gas.
[Stabilization Time]
Set the stabilization time.
[OK]
Applies the settings.
[Cancel]
Cancels the settings.