Checking the Instrument Status

In the [ICP-MSICP-QQQ MassHunter] window, the Instrument Status Pane gives a real-time display that lets you check the status of the instrument.

InstrumentStatusPane.png 

 

To display the pane, select [Panes] - [Instrument Status] from the [Show] group on the [View] tab.

 

The instrument status pane can show the values of 5 meters at the same time. To change which meters are displayed in the Instrument Status Pane, complete the following steps:

  1. Click [Meter] from the [Meters] group on the [View] tab.

    The [Select Meter] dialog box appears.

    SelectMeter_79.png

    [Select Meter] dialog box (For 7900)

  2.  

    The term "Carrier Gas" has been changed to "Nebulizer Gas" in MassHunter 4.4.

     

  3. Mark the check boxes of the meters you want to display.

    Up to 5 meters can be selected.

  4. Click [OK].

    The values of the selected meters are displayed in the Instrument Status Pane.

Meter Control Panel Values (For 7700)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

0.3 ~ 5

250 ~ 490

(Use Foreline Pump)

Pa

Analyzer Press

 

1x10-5 ~ 6x10-4

1x10-4 ~ 2x10-3

Pa

TMP Revolution

---

65 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 45

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

---

°C

Forward Power

0

0

700 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 500

kPa

OP Gas Tank Press

0

0

 

kPa

Ar Gas Tank Press

700

700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.8 ~ 1.3

L/min

MU/Dil. Gas

0

0

0 ~ 1.0

L/min

Optional Gas

0

0

 

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0 ~ 1.0

L/min

H2 Gas

0

0

3 ~ 5

mL/min

He Gas

0

0

3 ~ 5

mL/min

3rd Cell Gas

0

0

3 ~ 5

mL/min

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

Meter Control Panel Values (For 8800)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

0.3 ~ 5

200 ~ 490

Pa

Analyzer Press

 

1x10-5 ~ 7x10-4

1x10-4 ~ 2x10-3

Pa

TMP1 Revolution

0

80 ~ 100

95 ~ 100

%

TMP2 Revolution

0

80 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 40

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

20 ~ 50

°C

Forward Power

0

0

600 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 300

kPa

OP Gas Tank Press

0

0

10 ~ 100

kPa

Ar Gas Tank Press

700

700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.8 ~ 1.3

L/min

MU/Dil. Gas

0

0

0 ~ 1.0

L/min

Optional Gas

0

0

5 ~ 20

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0.9 ~ 1.5

L/min

H2 Gas

0

0

5 ~ 8

mL/min

He Gas

0

0

4 ~ 5, 7 ~ 10

mL/min

3rd Cell Gas

0

0

20 ~ 40

%

4th Cell Gas

0

0

20 ~ 40

%

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

 

3rd Cell Gas: Helium content of gas mix must be 90% or more; For example: NH3/He mix, He content must be equal to or greater than 90%; Even if introducing a non-corrosive non-flammable gas to the 3rd cell, it still must be diluted with 90% or more Helium.

4th Cell Gas: Use only non-flammable non-corrosive gas.

For important information and precautions on the handling and use of compressed gases (including oxygen, hydrogen and ammonia) refer to the Hardware Maintenance Manual, Ammonia Safety Guide, Hydrogen Safety Guide and Site Preparation Guide.

 

Meter Control Panel Values (For 7900)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

2 ~ 8

150 ~ 300

(Use Foreline Pump)

Pa

Analyzer Press

 

1x10-5 ~ 5x10-4

7x10-5 ~ 1x10-3

Pa

TMP Revolution

---

95 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 40

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

---

°C

Forward Power

0

0

700 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 500

kPa

OP Gas Tank Press

0

0

 

kPa

Ar Gas Tank Press

500 ~ 700

500 ~ 700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.7 ~ 1.2

L/min

MU/Dil. Gas

0

0

0 ~ 1.0

L/min

Optional Gas

0

0

0 ~ 40

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0.8 ~ 1.0

L/min

H2 Gas

0

0

3 ~ 5

mL/min

He Gas

0

0

4 ~ 6

mL/min

3rd Cell Gas

0

0

10 ~ 50

%

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

AVS Pump Speed

0

0

0 ~ 100

%

Valve Position

Load/Inject

Load/Inject

Load/Inject

-

Tune/ISTD Valve

Tune/ISTD

Tune/ISTD

Tune/ISTD

-

Meter Control Panel Values (For 7800)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

0.3 ~ 5

220 ~ 490

(Use Foreline Pump)

Pa

Analyzer Press

 

1x10-5 ~ 5x10-4

7x10-5 ~ 1x10-3

Pa

TMP Revolution

---

95 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 40

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

---

°C

Forward Power

0

0

700 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 500

kPa

OP Gas Tank Press

0

0

 

kPa

Ar Gas Tank Press

500 ~ 700

500 ~ 700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.7 ~ 1.2

L/min

MU/Dil. Gas

0

0

0 ~ 1.0

L/min

Optional Gas

0

0

0 ~ 40

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0.8 ~ 1.0

L/min

H2 Gas

0

0

3 ~ 5

mL/min

He Gas

0

0

4 ~ 6

mL/min

3rd Cell Gas

0

0

10 ~ 50

%

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

AVS Pump Speed

0

0

0 ~ 100

%

Valve Position

Load/Inject

Load/Inject

Load/Inject

-

Tune/ISTD Valve

Tune/ISTD

Tune/ISTD

Tune/ISTD

-

Meter Control Panel Values (For 8900)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

0.3 ~ 5

200 ~ 490

Pa

Analyzer Press

 

1x10-5 ~ 7x10-4

1x10-4 ~ 2x10-3

Pa

TMP1 Revolution

0

80 ~ 100

95 ~ 100

%

TMP2 Revolution

0

80 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 40

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

20 ~ 50

°C

Forward Power

0

0

600 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 300

kPa

OP Gas Tank Press

0

0

10 ~ 100

kPa

Ar Gas Tank Press

700

700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.8 ~ 1.3

L/min

MU/Dil. Gas

0

0

0 ~ 1.0

L/min

Optional Gas

0

0

5 ~ 20

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0.9 ~ 1.5

L/min

H2 Gas

0

0

5 ~ 8

mL/min

He Gas

0

0

4 ~ 5, 7 ~ 10

mL/min

3rd Cell Gas

0

0

20 ~ 40

%

4th Cell Gas

0

0

20 ~ 40

%

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

AVS Pump Speed

0

0

0 ~ 100

%

Valve Position

Load/Inject

Load/Inject

Load/Inject

-

Tune/ISTD Valve

Tune/ISTD

Tune/ISTD

Tune/ISTD

-

 

3rd Cell Gas: Helium content of gas mix must be 90% or more; For example: NH3/He mix, He content must be equal to or greater than 90%; Even if introducing a non-corrosive non-flammable gas to the 3rd cell, it still must be diluted with 90% or more Helium.

4th Cell Gas: Use only non-flammable non-corrosive gas.

For important information and precautions on the handling and use of compressed gases (including oxygen, hydrogen and ammonia) refer to the Hardware Maintenance Manual, Ammonia Safety Guide, Hydrogen Safety Guide and Site Preparation Guide.

 

Meter Control Panel Values (For 7850)

Meter

Typical Ranges

Units

Shutdown Mode

Standby Mode

Analysis Mode

IF/BK Press

---

0.3 ~ 5

220 ~ 490

(Use Foreline Pump)

Pa

Analyzer Press

 

1x10-5 ~ 5x10-4

7x10-5 ~ 1x10-3

Pa

TMP Revolution

---

95 ~ 100

95 ~ 100

%

Water RF/WC/IF

0.0

0.0

1 ~ 2

L/min

Water Temp

15 ~ 25

15 ~ 25

15 ~ 40

°C

Inlet Temp

Room Temp.

Room Temp.

15 ~ 35

°C

Internal Temp

Room Temp.

Room Temp.

---

°C

Forward Power

0

0

700 ~ 1600

W

Reflected Power

0

0

< 20

W

Plasma Freq.

---

---

26.0 ~ 28.0

MHz

Nebulizer Gas (BP)

0

0

200 ~ 500

kPa

OP Gas Tank Press

0

0

 

kPa

Ar Gas Tank Press

500 ~ 700

500 ~ 700

500 ~ 700

kPa

Nebulizer Gas

0

0

0.7 ~ 1.2

L/min

Optional Gas

0

0

0 ~ 40

%

Plasma Gas

0

0

15

L/min

Aux Gas

0

0

0.8 ~ 1.0

L/min

H2 Gas

0

0

3 ~ 5

mL/min

He Gas

0

0

4 ~ 6

mL/min

S/C Temp (L)

Room Temp.

Room Temp.

2

°C

AVS Pump Speed

0

0

0 ~ 100

%

Valve Position

Load/Inject

Load/Inject

Load/Inject

-

Tune/ISTD Valve

Tune/ISTD

Tune/ISTD

Tune/ISTD

-