In the [ICP-MSICP-QQQ MassHunter] window, the Instrument Status Pane gives a real-time display that lets you check the status of the instrument.
To display the pane, select [Panes] - [Instrument Status] from the [Show] group on the [View] tab.
The instrument status pane can show the values of 5 meters at the same time. To change which meters are displayed in the Instrument Status Pane, complete the following steps:
The [Select Meter] dialog box appears.
[Select Meter] dialog box (For 7900)
The term "Carrier Gas" has been changed to "Nebulizer Gas" in MassHunter 4.4.
Up to 5 meters can be selected.
The values of the selected meters are displayed in the Instrument Status Pane.
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
0.3 ~ 5 |
250 ~ 490 (Use Foreline Pump) |
Pa |
Analyzer Press |
1x10-5 ~ 6x10-4 |
1x10-4 ~ 2x10-3 |
Pa |
|
TMP Revolution |
--- |
65 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 45 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
--- |
°C |
Forward Power |
0 |
0 |
700 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 500 |
kPa |
OP Gas Tank Press |
0 |
0 |
kPa |
|
Ar Gas Tank Press |
700 |
700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.8 ~ 1.3 |
L/min |
MU/Dil. Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
Optional Gas |
0 |
0 |
% |
|
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
H2 Gas |
0 |
0 |
3 ~ 5 |
mL/min |
He Gas |
0 |
0 |
3 ~ 5 |
mL/min |
0 |
0 |
3 ~ 5 |
mL/min |
|
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
0.3 ~ 5 |
200 ~ 490 |
Pa |
Analyzer Press |
1x10-5 ~ 7x10-4 |
1x10-4 ~ 2x10-3 |
Pa |
|
TMP1 Revolution |
0 |
80 ~ 100 |
95 ~ 100 |
% |
TMP2 Revolution |
0 |
80 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 40 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
20 ~ 50 |
°C |
Forward Power |
0 |
0 |
600 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 300 |
kPa |
OP Gas Tank Press |
0 |
0 |
10 ~ 100 |
kPa |
Ar Gas Tank Press |
700 |
700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.8 ~ 1.3 |
L/min |
MU/Dil. Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
Optional Gas |
0 |
0 |
5 ~ 20 |
% |
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0.9 ~ 1.5 |
L/min |
H2 Gas |
0 |
0 |
5 ~ 8 |
mL/min |
He Gas |
0 |
0 |
4 ~ 5, 7 ~ 10 |
mL/min |
0 |
0 |
20 ~ 40 |
% |
|
4th Cell Gas |
0 |
0 |
20 ~ 40 |
% |
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
3rd Cell Gas: Helium content of gas mix must be 90% or more; For example: NH3/He mix, He content must be equal to or greater than 90%; Even if introducing a non-corrosive non-flammable gas to the 3rd cell, it still must be diluted with 90% or more Helium.
4th Cell Gas: Use only non-flammable non-corrosive gas.
For important information and precautions on the handling and use of compressed gases (including oxygen, hydrogen and ammonia) refer to the Hardware Maintenance Manual, Ammonia Safety Guide, Hydrogen Safety Guide and Site Preparation Guide.
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
2 ~ 8 |
150 ~ 300 (Use Foreline Pump) |
Pa |
Analyzer Press |
1x10-5 ~ 5x10-4 |
7x10-5 ~ 1x10-3 |
Pa |
|
TMP Revolution |
--- |
95 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 40 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
--- |
°C |
Forward Power |
0 |
0 |
700 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 500 |
kPa |
OP Gas Tank Press |
0 |
0 |
kPa |
|
Ar Gas Tank Press |
500 ~ 700 |
500 ~ 700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.7 ~ 1.2 |
L/min |
MU/Dil. Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
Optional Gas |
0 |
0 |
0 ~ 40 |
% |
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0.8 ~ 1.0 |
L/min |
H2 Gas |
0 |
0 |
3 ~ 5 |
mL/min |
He Gas |
0 |
0 |
4 ~ 6 |
mL/min |
0 |
0 |
10 ~ 50 |
% |
|
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
AVS Pump Speed |
0 |
0 |
0 ~ 100 |
% |
Valve Position |
Load/Inject |
Load/Inject |
Load/Inject |
- |
Tune/ISTD Valve |
Tune/ISTD |
Tune/ISTD |
Tune/ISTD |
- |
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
0.3 ~ 5 |
220 ~ 490 (Use Foreline Pump) |
Pa |
Analyzer Press |
1x10-5 ~ 5x10-4 |
7x10-5 ~ 1x10-3 |
Pa |
|
TMP Revolution |
--- |
95 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 40 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
--- |
°C |
Forward Power |
0 |
0 |
700 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 500 |
kPa |
OP Gas Tank Press |
0 |
0 |
kPa |
|
Ar Gas Tank Press |
500 ~ 700 |
500 ~ 700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.7 ~ 1.2 |
L/min |
MU/Dil. Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
Optional Gas |
0 |
0 |
0 ~ 40 |
% |
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0.8 ~ 1.0 |
L/min |
H2 Gas |
0 |
0 |
3 ~ 5 |
mL/min |
He Gas |
0 |
0 |
4 ~ 6 |
mL/min |
0 |
0 |
10 ~ 50 |
% |
|
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
AVS Pump Speed |
0 |
0 |
0 ~ 100 |
% |
Valve Position |
Load/Inject |
Load/Inject |
Load/Inject |
- |
Tune/ISTD Valve |
Tune/ISTD |
Tune/ISTD |
Tune/ISTD |
- |
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
0.3 ~ 5 |
200 ~ 490 |
Pa |
Analyzer Press |
1x10-5 ~ 7x10-4 |
1x10-4 ~ 2x10-3 |
Pa |
|
TMP1 Revolution |
0 |
80 ~ 100 |
95 ~ 100 |
% |
TMP2 Revolution |
0 |
80 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 40 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
20 ~ 50 |
°C |
Forward Power |
0 |
0 |
600 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 300 |
kPa |
OP Gas Tank Press |
0 |
0 |
10 ~ 100 |
kPa |
Ar Gas Tank Press |
700 |
700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.8 ~ 1.3 |
L/min |
MU/Dil. Gas |
0 |
0 |
0 ~ 1.0 |
L/min |
Optional Gas |
0 |
0 |
5 ~ 20 |
% |
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0.9 ~ 1.5 |
L/min |
H2 Gas |
0 |
0 |
5 ~ 8 |
mL/min |
He Gas |
0 |
0 |
4 ~ 5, 7 ~ 10 |
mL/min |
0 |
0 |
20 ~ 40 |
% |
|
4th Cell Gas |
0 |
0 |
20 ~ 40 |
% |
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
AVS Pump Speed |
0 |
0 |
0 ~ 100 |
% |
Valve Position |
Load/Inject |
Load/Inject |
Load/Inject |
- |
Tune/ISTD Valve |
Tune/ISTD |
Tune/ISTD |
Tune/ISTD |
- |
3rd Cell Gas: Helium content of gas mix must be 90% or more; For example: NH3/He mix, He content must be equal to or greater than 90%; Even if introducing a non-corrosive non-flammable gas to the 3rd cell, it still must be diluted with 90% or more Helium.
4th Cell Gas: Use only non-flammable non-corrosive gas.
For important information and precautions on the handling and use of compressed gases (including oxygen, hydrogen and ammonia) refer to the Hardware Maintenance Manual, Ammonia Safety Guide, Hydrogen Safety Guide and Site Preparation Guide.
Meter |
Typical Ranges |
Units |
||
---|---|---|---|---|
Shutdown Mode |
Standby Mode |
|||
IF/BK Press |
--- |
0.3 ~ 5 |
220 ~ 490 (Use Foreline Pump) |
Pa |
Analyzer Press |
1x10-5 ~ 5x10-4 |
7x10-5 ~ 1x10-3 |
Pa |
|
TMP Revolution |
--- |
95 ~ 100 |
95 ~ 100 |
% |
Water RF/WC/IF |
0.0 |
0.0 |
1 ~ 2 |
L/min |
Water Temp |
15 ~ 25 |
15 ~ 25 |
15 ~ 40 |
°C |
Inlet Temp |
Room Temp. |
Room Temp. |
15 ~ 35 |
°C |
Internal Temp |
Room Temp. |
Room Temp. |
--- |
°C |
Forward Power |
0 |
0 |
700 ~ 1600 |
W |
Reflected Power |
0 |
0 |
< 20 |
W |
Plasma Freq. |
--- |
--- |
26.0 ~ 28.0 |
MHz |
Nebulizer Gas (BP) |
0 |
0 |
200 ~ 500 |
kPa |
OP Gas Tank Press |
0 |
0 |
kPa |
|
Ar Gas Tank Press |
500 ~ 700 |
500 ~ 700 |
500 ~ 700 |
kPa |
Nebulizer Gas |
0 |
0 |
0.7 ~ 1.2 |
L/min |
Optional Gas |
0 |
0 |
0 ~ 40 |
% |
Plasma Gas |
0 |
0 |
15 |
L/min |
Aux Gas |
0 |
0 |
0.8 ~ 1.0 |
L/min |
H2 Gas |
0 |
0 |
3 ~ 5 |
mL/min |
He Gas |
0 |
0 |
4 ~ 6 |
mL/min |
S/C Temp (L) |
Room Temp. |
Room Temp. |
2 |
°C |
AVS Pump Speed |
0 |
0 |
0 ~ 100 |
% |
Valve Position |
Load/Inject |
Load/Inject |
Load/Inject |
- |
Tune/ISTD Valve |
Tune/ISTD |
Tune/ISTD |
Tune/ISTD |
- |