Follow the steps below to set up the maintenance position for exchanging the sample probe:
Sample Introduction from the [Settings] dialog box is displayed.
This item is only displayed when SPS 4 is selected as the [Sample Introduction] and [Use Autosampler] is selected.
The [Maintenance] dialog box is displayed.
Select this item to set the maintenance position to the bottle-5 position on the 5 wall standards rack (Z-axis position is 0 mm).
Select this item to set the maintenance position specified by the X/Y/Z-axes values.
At this time, you can exchange the sample probe from the position.
If you need to initialize SPS 4 after replacing the sample probe, select [Initialize Autosampler at closing dialog box], and then click [Close].