Sets the conditions that are used to determine outliers when FullQuant Analysis is selected in the Data Analysis Method table.
Drag to move the toolbar.
Displays
the settings for the previous basic method task.
[Edit
QC Level Caption] dialog box will be displayed. Set the QC level
captions.
This section describes the commands that are available in the context menu. The commands that are not listed below have the same function as their counterparts on the toolbars and the Ribbon.
Opens the [Export Table] (Entire Table) dialog box where you can export the entire table. This command is available when a pane that includes a table is selected.
Opens the [Export Table] (Area) dialog box where you can export the data selected in the table. This command is available when a range of cells in the table is selected.
Copies the selected data to the Clipboard. When a tabular pane is selected, the selected row/column is copied.
Pastes the data copied with the Copy command.
Opens the [Page Setup] dialog box where you can configure the page settings for printing.
Opens the [Print] dialog box where you can print the contents of the selected pane.
Opens the [Print Preview] dialog box where you can display the print preview.
Sets the conditions that are used to determine outliers.
Displays the Outlier type.
Outlier in the correlation coefficient of the calibration curve
Outlier in the Relative Standard Error (%) of the calibration curve
Outlier in the Relative Error (%) of the calibration curve
Outlier for ISTD recovery by comparison with CalBlk
Outlier for Spike recovery by comparison with SpikeRef
Outlier for the QC sample concentration stability of the QC1 sample
Outlier for the QC sample concentration stability of the QC2 sample
Outlier for the QC sample concentration stability of the QC3 sample
Outlier for the QC sample concentration stability of the QC4 sample
Outlier for the QC sample concentration stability of the QC5 sample
Outlier for the count RSD % (relative standard deviation). For the reference, select [Check Always] or enter a cps value directly.
Outlier for the blank concentration stability of the blank sample
Outlier outside the concentration range of the calibration curve
Select the minimum value to be determined as an Outlier
Select the maximum value to be determined as an Outlier
Displays the reference sample type for Outlier judgment. Input the count (cps) for CountRSD
An action can be set when an outlier occurs during measurement.
The terms "AOF", "Action on Failure", and "Action Failure" have been changed to "Error Action" in MassHunter 4.4.
No actions will be made even if an outlier occurs (default).
Stop the sequence
Analyze the next sample after analyzing a [BLANK] block. (Continue the sequence even if the blank fails.)
Analyze the same sample after regenerating [CALIB]. (Stop the sequence for the second error.)
When using prepFAST as an autosampler or ADS 2 as a sample introduction, you can select this as the outlier for [ISTD Recovery %] and [Out of Calibration Curve Concentration Range %]. If a sample analysis fails, the sample is diluted and analyzed again. If the sample is reanalyzed twice and the third analysis fails, the next sample is analyzed.