Click [Configure Ignition Sequence] on the Startup
Pane toolbar, and then select [User Tune]. Also turn on [Set as
Global Tune] in the User
Tune Pane toolbar.
Setup a No Gas tune mode in User tune. Tune it as usual, using
peripump as a sample introduction system. Because the standard LC
flow rate is higher than the usual peripump speed, set the peripump
speed higher than under normal conditions. 1 ml/min ≒ 0.3 rps.
Select No Gas mode as the Set Base Tune for Startup. For an s-model,
select <None> as the Base Cool Tune Mode.
Turn on "Custom Settings" in Startup tasks. Set the masses
included in your P/A factor solution as the P/A factor adjustment
parameters.
Select [PeriPump] as the Sample
Introduction. Do not connect LC tubing to ICP-MS; instead keep
the sample introduction tubing using the peripump.
Ignite plasma and run the Startup.
Turn on the LC pump to stabilize, but do not introduce the solution
from LC to ICP-MS.
When the Startup tasks are operating except for the P/A factor, introduce
1ppb tune solution to the peripump.
When the Startup tasks are operating for the P/A factor, introduce
the P/A factor adjustment solution to the peripump.
Stop the LC pump and connect its tubing to the ICP-MS, and then
restart the LC pump.