MassHunter Workstation categorizes data by sample type for data analysis.
The uses for each sample type are defined as follows.
Sample type |
Full Name |
Use |
---|---|---|
Bkgnd |
Background |
Set for background subtraction blanks. The data in this sample is subtracted from each sample data acquired. |
BlkVrfy |
Blank Verify |
Set for blanks for blank concentration stability. Lets you verify calibration curve by measuring this sample regularly. The blank is quantified and becomes an outlier if the result exceeds the standard. |
CalBlk |
Calibration Blank |
Set for blanks for calibration curves and used as the reference for ISTD Stability Graph. Samples used as reference for the ISTD Stability Graph can be changed in the Batch Table. Right-click on the row of the sample to set as a reference and select [Set as ISTD Reference File] from the context menu. |
CalStd |
Calibration Standard |
Set for standard solution samples for calibration curves. Set the calibration curve level in the Level column of the Batch Table. |
DilIS |
Dilution ISTD |
Set for standard solution samples for simultaneous correction when using ISIS 2 Autodilution. However, new data acquisition is not possible. |
DilStd |
Dilution Standard |
Set for standard solution samples for periodic correction when using ISIS 2 Autodilution. However, new data acquisition is not possible. |
DriftChk |
Drift Check |
Set to standard solution samples when executing “Reslope-NextSmpl” in the optional Intelligent Sequence application. |
FQBlk |
Full Quant Blank |
Set for blanks for blank concentration subtraction in FullQuant Analysis. The concentration of this sample is subtracted from each sample data that is acquired. |
CICSpike |
CICSpike |
Set for spiked samples in the optional chromatographic analysis. |
IsoStd |
Isotope Analysis Standard |
Set for standard solution samples for mass bias correction in Isotope Ratio Analysis and Isotope Dilution Analysis. |
QC1 to 5 |
Quality Check 1 to 5 |
Set for QC samples used for FullQuant Outliers and QC Sample Stability Graphs. There are three types of QC samples: If you have set to show outliers from QC1 to 5 in the [FullQuant Outliers] dialog box, you must enter the concentration for QC1 to 5 in the Calibration pane of the Analysis Method. |
Sample |
Sample |
Set for unknown samples to be analyzed. |
SQBlk |
Semi Quant Blank |
Set for blanks for blank concentration subtraction in SemiQuant Analysis. The concentration of this sample is subtracted from each sample data that is acquired. |
SQISTD |
Semi Quant ISTD |
Set for standard solution samples for internal standard correction (Auto Add Mode) in SemiQuant Analysis. |
SQStd |
Semi Quant Standard |
Set for standard solution samples for SemiQuant factor correction in SemiQuant Analysis. |
RM |
Reference Material |
Set for the reference material in the Single Particle Analysis. |
Spike |
Spike Sample |
Set for the spiked sample for recovery. |
SpikeRef |
Spike Reference Sample |
Set for the Spike Reference sample for recovery. |
IonicBlk |
Analyte response factor calibration blank |
Blank for the response factor calibration of Set for Single Particle Analysis. |
IonicStd (AN) |
Analyte response factor calibration standard |
Standard sample for the target element response factor calibration of Set for Single Particle Analysis. |
IonicStd (RM) |
Reference material response factor calibration standard |
Standard sample for the reference element response factor calibration of Set for Single Particle Analysis. |
Sample type |
Full Name |
Use |
---|---|---|
Bkgnd |
Background |
Set for background subtraction blanks. The data in this sample is subtracted from each sample data acquired. |
BlkVrfy |
Blank Verify |
Set for blanks for blank concentration stability. Lets you verify calibration curve by measuring this sample regularly. The blank is quantified and becomes an outlier if the result exceeds the standard. |
CalBlk |
Calibration Blank |
Set for blanks for calibration curves and used as the reference for ISTD Stability Graph. Samples used as reference for the ISTD Stability Graph can be changed in the Batch Table. Right-click on the row of the sample to set as a reference and select [Set as ISTD Reference File] from the context menu. |
CalStd |
Calibration Standard |
Set for standard solution samples for calibration curves. Set the calibration curve level in the Level column of the Batch Table. |
DriftChk |
Drift Check |
Set to standard solution samples when executing “Reslope-NextSmpl” in the optional Intelligent Sequence application. |
FQBlk |
Full Quant Blank |
Set for blanks for blank concentration subtraction in FullQuant Analysis. The concentration of this sample is subtracted from each sample data that is acquired. |
CICSpike |
CICSpike |
Set for spiked samples in the optional chromatographic analysis. |
IsoStd |
Isotope Analysis Standard |
Set for standard solution samples for mass bias correction in Isotope Ratio Analysis and Isotope Dilution Analysis. |
QC1 to 5 |
Quality Check 1 to 5 |
Set for QC samples used for FullQuant Outliers and QC Sample Stability Graphs. There are three types of QC samples: If you have set to show outliers from QC1 to 5 in the [FullQuant Outliers] dialog box, you must enter the concentration for QC1 to 5 in the Calibration pane of the Analysis Method. |
Sample |
Sample |
Set for unknown samples to be analyzed. |
SQBlk |
Semi Quant Blank |
Set for blanks for blank concentration subtraction in SemiQuant Analysis. The concentration of this sample is subtracted from each sample data that is acquired. |
SQISTD |
Semi Quant ISTD |
Set for standard solution samples for internal standard correction (Auto Add Mode) in SemiQuant Analysis. |
SQStd |
Semi Quant Standard |
Set for standard solution samples for SemiQuant factor correction in SemiQuant Analysis. |
RM |
Reference Material |
Set for the reference material in the Single Particle Analysis. |
Spike |
Spike Sample |
Set for the spiked sample for recovery. |
SpikeRef |
Spike Reference Sample |
Set for the Spike Reference sample for recovery. |
IonicBlk |
Analyte response factor calibration blank |
Blank for the response factor calibration of Set for Single Particle Analysis. |
IonicStd (AN) |
Analyte response factor calibration standard |
Standard sample for the target element response factor calibration of Set for Single Particle Analysis. |
IonicStd (RM) |
Reference material response factor calibration standard |
Standard sample for the reference element response factor calibration of Set for Single Particle Analysis. |