This dialog box lets you select additional operations to enable for the batch execution.
Tune check before batch
This item is displayed when [Tune Check function in Batch] is selected in the [Options] category in the [Settings] dialog box. Mark the check box to do a tuning check before you run the batch.
In addition, sets actions to take when a tuning failure occurs.
In previous versions, the parameters for Performance Report in Hardware Settings are used for auto tune and tune report. These are always used for parameters in the batch.
Print batch log at end
Selects to print the batch log.
Tune mode stabilization time
This item is displayed when [TRA] is selected as the [Acq Mode]. Select this to set the [Stabilization Time] for the TRA measurement.
Rinse after batch
Mark the check box to do the rinse at the end of the batch.
[Set Rinse after Batch]
Lets you configure the rinse processes on the [Set Rinse after Batch] dialog box when you select "Rinse after Batch".
[Calibration Import]
You can import calibration data (standard data and analysis methods) from another batch.
[Previous Batch]: Imports the file of the previous batch.
[Select Batch]: Select a batch folder to be imported.
For Single Particle Analysis, the item name changes to [Import Reference Sample and Calibration Samples].
[Enhanced Stability Mode]
This item is displayed when [Enable Enhanced Stability Tune Mode] is selected in [Options] category in the [Settings] dialog box.
This lists “Typical Environmental (ICS-A)” or “Typical Material Matrix (1% Cu)”. The parameters are used for Enhanced Stability Mode. If “None” is selected, Enhanced Stability Mode is off. The parameters should be configured in the [Conditioning settings] dialog box.
For details, refer to the "Enhanced Stability Mode (ESM)".
[Recovery from Over Range]
This item is enabled in TRA mode.
When its signal level exceeds the analog mode limit, the detector closes the analog gate and no measurement is made for the sample. When the signal level drops below the analog mode limit, it opens the analog gate and resumes measurement. This parameter defines the time interval for checking the signal level. It checks the signal level three times in the specified period. The available setting range is from 3 to 6000 seconds.
[Set Pre and Post Script]
On the [Set Pre and Post Script] dialog box, configure the scripts to run before and after the batch.
[OK]
Applies the settings.
[Cancel]
Cancels the settings.