Copy or remove the current tune mode.
Tune Mode
Displays the current tune mode.
[OK]
Applies the settings.
[Cancel]
Cancels the settings.
[Duplicate]
Duplicates the selected tune mode.
You can set up to 16 tune modes.
For Rapid Multi-Element Nanoparticle Analysis, you can set 17 or more tune modes. To use this feature, turn on [sNP Extended Functions] from the [Add-Ins] group on the [Tools] tab, and turn on [Extended Acquisition] from the [sNP Extended Functions] group on the [sNP Extension] tab displayed.
When you duplicate a gas tune mode to use for a different tune mode (for example: duplicating an H2 tune to use as the basis for an He tune), use only semi-autotune.
To use Semi Autotune, select the [Auto/Semi Auto Tune before Batch] check box in the Setup Pane, and then select Custom Tune in the [Configure Tune Way] dialog box.
When you duplicate a gas tune mode to use for a different tune mode (for example: duplicate an O2 tune to use as the basis for an NH3 tune), use only semi autotune.
To use Semi Autotune, select the [Auto/Semi Auto Tune before Batch] check box in the Setup Pane, and then select Custom Tune in the [Configure Tune Way] dialog box.
[Delete]
Deletes the selected tune mode.