Organic Solvent Analysis Workflow
The following describes the differences in organic solvent analyses
on the instrument
based on the general elemental analysis workflow on MassHunter Workstation.
Operation Procedure Overview
Usually, start the operation of the instrument in Standby mode. The
procedure for putting the instrument in Standby mode when the power is
off is the same as the general procedure. In addition to the standard
preparation, make sure that 20% oxygen/argon is supplied to the optional
gas for the instrument.
Startup for Organic Solvent Analysis
In a startup for analysis of organic solvents, as in the case of aqueous
solutions, the torch axis and mass axis are adjusted after plasma ignition
to optimize the basic performance of the instrument. Isopropyl alcohol
is used when the plasma is ignited.
Note that EM adjustment
is not performed for organic solvent analyses, but it is for aqueous solutions.
For more information on startup settings, refer to the startup for general
aqueous solutions in Executing Startup.
Conditions for Performing
a Startup
In organic solvent analyses, the operation procedure varies depending
on whether a Startup is performed or not. Refer to Executing Startup for
more information. To perform Startup, the instrument must meet the following
conditions:
Part numbers are subject to change.
A startup cannot be performed for other conditions (e.g., s-lens
is used). In such a case, refer to "When
No Startup Is Performed" below and also "Operation
Procedure When No Startup Is Performed."
When a Startup Is Performed
In this case, perform a startup after introducing isopropyl alcohol
and igniting the plasma. Next, load a batch suitable for the solvent to
be analyzed and perform lens tuning. After that, introduce the sample
and start the analysis. You can adjust plasma conditions manually.
For details on the operation, refer to "Operation
Procedure When a Startup Is Performed"
When No Startup Is Performed
In this case, perform tuning manually after introducing isopropyl alcohol
and igniting the plasma. Perform lens tuning and start the sample analysis.
You can adjust plasma conditions manually.
For details on the operation, refer to "Operation
Procedure When No Startup Is Performed."
Solutions to Be Prepared
When you perform a Startup, prepare the following solutions in advance:
- Rinse solution
Before starting the analysis, use isopropyl alcohol. During the
analysis, use the target solvent.
- Tuning solution
Prepare a tuning solution by diluting a standard solution of an
aqueous solvent with isopropyl alcohol.
Solvent: Isopropyl alcohol. Li, Co, Y, Ce, and Tl at 10 mg/L
Operation
Procedure When a Startup Is Performed
Start MassHunter Workstation
- Start the MassHunter Workstation software.
Configure Hardware
- Set an O ring for the connector of the sample tube used for the
nebulizer for organic solvents (if the connector is used), a sample
introduction tube, and a torch. Set a drain tube for organic solvents
that is suitable for the target solvent. (For other conditions on
the instrument, refer to "Conditions
for Performing a Startup.")
- The peripump is used only for drainage and is not used for introducing
liquids to the nebulizer.
- Configure the hardware settings on the MassHunter Workstation.
On [System] - [Hardware] in the [Settings]
dialog box, specify [Organic Solvent] for Ignition Mode, [MicroMist]
for Nebulizer, and [PeriPump] for Sample Introduction. Select [x-Lens]
for Ion Lenses.
Configure User Tune
Follow the steps below to configure User Tune.
- Click [User Tune Configuration] from the [Startup] group in the
Task Navigator.
To display the [Startup] group, click [Configuration]
from the [Startup] group on the [Home] tab.
The User Tune Pane
appears.
- Select [IPA] from the [Organic Solvent Parameters] pull-down menu.
- Select [No Gas] for [Set Base Tune for Startup].
- Click [Startup Configuration] from the [Startup] group in the Task Navigator.
- Click [Configure Ignition Sequence] on the Startup
Pane toolbar.
The [Configure
Ignition Sequence] dialog box is displayed.
- Select [User Tune], and then click [OK].
Configure Startup
- After enabling Custom Settings (click [Select Custom Settings]
on the Startup Pane taskbar), configure as follows. Do not configure
Plasma Correction, EM or
P/A factor.
- In the standard settings, the rinse port is specified to be used.
For some autosamplers, no organic solvent can be used as a rinse solution
for the rinse port. In such a case, change the setting of [Set] in
[PeriPump] so that the rinse port is not used. Be sure to allow sufficient
Sample Uptake and Stabilization time. For details, refer to the instruction
manual for each autosampler.
Turn Plasma ON / Execute Startup
- Click the [Plasma] from the [Startup] group in the [Home] tab.
When the message "Do you want to perform a startup after plasma
ignition?" is displayed, select [Yes].
- After Startup is complete, introduce the tuning solution.
Replace Solvent
- If the solvent used for the analysis is not isopropyl alcohol,
introduce the target solvent.
- Select the rinse solution suitable for the target solvent.
Edit Batch
- Specify the conditions for the acquisition method, data analysis
method, and sample list, excluding the conditions for tuning.
- Subsequent steps are the same as in the general operation for aqueous
solution analyses.
Execute Queue
- Add the created batch to the queue and start the acquisition.
Check Analysis Results
- The analysis results can be viewed on the panes in the ICP-MSICP-QQQ Data
Analysis window.
Generate Reports
- Generate a sample report or quick batch report.
Save Analysis Results
Turn Plasma OFF
Exit MassHunter Workstation
- Exit the MassHunter Workstation software.
Operation
Procedure When No Startup Is Performed
Start MassHunter Workstation
- Start the MassHunter Workstation software.
Configure Hardware
- Select MicroMist Nebulizer or MicroFlow Nebulizer (PFA) as the
nebulizer. Set a sample introduction tube suitable for the sample,
a torch, a drain tube for organic solvents, and an O ring for the
connector of the sample tube used for the nebulizer for organic solvents
(if the connector is used). For more information on selecting a sample
introduction tube and torch, refer to Appendix A in the Hardware Maintenance
Manual.
- The peripump is used only for drainage and is not used for introducing
liquids to the nebulizer.
- Configure the hardware settings on the MassHunter Workstation.
On the Dashboard Pane,
select [Organic Solvent] for Ignition Mode and your nebulizer for
Nebulizer. Set [PeriPump] for Sample Introduction. Select the lens
to use for Lens.
Configure User Tune
Follow the steps below to configure User Tune.
- Click [User Tune Configuration] from the [Startup] group in the
Task Navigator.
To display the [Startup] group, click [Configuration]
from the [Startup] group on the [Home] tab.
The User Tune Pane
appears.
- Select [IPA] from the [Organic Solvent Parameters] pull-down menu.
- Select [No Gas] for [Set Base Tune for Startup].
- Click [Startup Configuration] from the [Startup] group in the Task Navigator.
- Click [Configure Ignition Sequence] on the Startup
Pane toolbar.
The [Configure
Ignition Sequence] dialog box is displayed.
- Select [User Tune], and then click [OK].
Configure Startup
- In this case, you do not need to configure a startup.
Turn Plasma ON
- Click the [Plasma] from the [Startup] group in the [Home] tab.
When the message "Do you want to perform a startup after plasma
ignition?" is displayed, select [No].
- In Analysis
Mode after ignition, you can adjust Torch Axis and Resolution/Axis
in the user tune. You can also use the Dashboard
Pane to adjust Torch Axis and Resolution/Axis.
- When you adjust them manually, introduce the tuning solution.
Replace Solvent
- If the solvent used for the analysis is not isopropyl alcohol,
introduce the target solvent.
- Select the rinse solution suitable for the target solvent.
Edit Batch
- Specify the conditions for the acquisition method, data analysis
method, and sample list, excluding the conditions for tuning.
- Subsequent steps are the same as in the case of When
a Startup Is Performed and the general operation for aqueous
solution analyses.
Execute Queue
- Add the created batch to the queue and start the acquisition.
Check Analysis Results
- The analysis results can be viewed on the panes in the ICP-MSICP-QQQ Data
Analysis window.
Generate Reports
- Generate a sample report or quick batch report.
Save Analysis Results
Turn Plasma OFF
Exit MassHunter Workstation
- Exit the MassHunter Workstation software.
Differences between Organic Solvents and Aqueous Solutions
When an organic solvent is used, the gas flow rate and control timing
used when the plasma is ignited are different from those of an aqueous
solution. This is because the volatility, the viscosity, and the amount
of the solvent introduced to the plasma vary depending on each organic
solvent.
After the plasma is ignited, set the tuning parameters used in Analysis
Mode so that they are suitable for organic solvents. The biggest
differences from the case of aqueous solutions are in the nebulizer
gas (also called carrier gas) flow rate, optional gas flow rate,
makeup gas flow rate, peripump, and spray chamber temperature.
These differences in parameters are due to the following factors:
- To prevent carbons from being deposited as soot on the interface
when the (organic solvent) sample is decomposed by plasma, oxygen
is introduced through the optional gas.
- If a large amount of the (organic solvent) sample is introduced,
the plasma may not be maintained. Also, a large amount of oxygen is
also required.
- Some (organic solvent) samples deteriorate a peripump tube. So,
such a sample is introduced by self-aspiration instead of using a
peripump.
- Because no peripump is used, the introduction amount of the sample
is controlled by changing the nebulizer gas flow rate and the inner
diameter of the sample introduction tube.
- The peripump is used only for draining liquids.
- To reduce the vapor pressure of the (organic solvent) sample, the
temperature of the spray chamber is set to -5ºC.
- In MassHunter Workstation, the above parameters are set in the
batches for organic solvents. The available batch types are IPA, Kerosene,
and Xylene, and the difficulty of maintaining the plasma increases
in this order. If you cannot find a batch with a relevant solvent
name, select the one whose difficulty of maintaining the plasma is
closest to that of your solvent.
When you select Xylene but the plasma cannot be maintained
or soot is deposited, replace the introduction tube with a thinner one
and reduce the introduction amount of the sample.