Chromatogram Pane Operations

This section explains the operation of the Chromatogram pane (Option), which is displayed when you analyze samples with the optional Chromatographic Application. For information about functions that are common to all panes, refer to “Common Pane Operations”.

 

Chromatogram Pane

Displaying Chromatograms

The Chromatogram pane (Option is displayed when you load data for which the Analysis Mode is set to “Chromatogram”. The chromatogram for the sample that is selected in the Batch Table is displayed in the Chromatogram pane (Option).

The Chromatogram pane (Option) has two rows. The upper row displays a chromatogram for the entire time range of an EIC (extracted ion chromatogram), and highlights the currently selected peak in purple. The lower row displays an expanded view of the analyte peak that is selected on the Batch table. In both rows, the X-axis represents the retention time in minutes, and the Y-axis represents the counts.

To display a different chromatogram in the Chromatogram pane (Option), click one of the following icons on the Chromatogram pane toolbar.

PreviousSample.png   Displays the chromatogram for the previous sample.

NextSample.png   Displays the chromatogram for the next sample.

PreviousElement.png   Selects the previous peak.

NextElement.png   Selects the next peak.

 

If the Chromatogram pane (Option) pane is not shown, click [Pane] from the [Show] group on the [View] tab. Also check that  on the toolbar is selected (when a toolbar icon is selected, it is highlighted in a darker shade). If not, click to select it.

 

Toolbar Functions

For the functions of the toolbar, refer to “Spectrum pane” help topic.

Adjusting the scales

The scale for the X-and Y-axes can be changed. For more information, refer to “Adjusting the scales” under “Common Graph Operations”.

Shifting the axes

The X-and Y-axes can be transposed. For more information, refer to “Shifting the axes” under “Common Graph Operations”.

Expanding a selected area

You can zoom in on a selected area of the chromatogram. For more information, refer to “Expanding a selected area” under “Common Graph Operations”.

Changing the display type

You can change the chromatogram display type by clicking on the toolbar icons.

image247.jpg   Displays or hides the tooltip display for the Count value.

image335.jpg   Displays or hides the peak baseline.

image337.jpg   Changes the display between TIC (total ion chromatogram) and EIC (extraction ion chromatogram). By default, EIC is displayed. Click and invert image33700554.jpg to display the TIC.

image365.jpg   Displays or hides the Target ion.

image367.jpg   Displays or hides the qualifier ions.

Displaying labels

Labels can be displayed in the upper-row chromatogram and the lower-row peak, as shown below.

LabelDisplay.png 

Label Display

To display the labels, follow these steps:

  1. Click the image341.jpg icon on the toolbar.

  2. The [Chromatogram Information] dialog box is displayed.

    Chromatogram_Information.png 

    [Chromatogram Information] Dialog Box

  3. Select the items to display.

    • [EIC/TIC]: Select the items to display in the upper chromatogram.
    • [Analyte]: Select the items to be displayed in the lower chromatogram.
  4. Click <OK>.

    The selected labels are shown in the chromatograms.

 

To hide all labels, clear all checks in the [Chromatogram Information] dialog box and click <OK>.

 

Displaying mass spectra

You can view the mass spectra for any time slice.

Place the cursor at a desired point on the chromatogram, then double-click to display the mass spectra for that particular point in time.

To display the mass spectrum for another peak, click PreviousElement.png NextElement.png on the toolbar in the Spectrum pane.

To return to the chromatogram display, double-click on the pane.

Adding comments

Comments can be added to the chromatograms.

Adding a comment

To add a comment, follow these steps:

  1. Click the image259.jpg icon on the toolbar.

    The icon is selected, and the cursor changes to image261.jpg.

  2. Click at the location to add the comment.

    The [Add comment] dialog box is displayed.

    image263.png 

    [Add comment] Dialog Box

  3. Type in your comment.
  4. Click <Add>.

    The comment is shown at the specified point on the chromatogram.

  5. When finished, click image25900555.jpg again.

 

To save an annotation in the batch, select the [Annotate] - [Save Annotation in Data] from the context menu.

 

Deleting comments

To delete comments, follow these steps:

  1. Click the image25900556.jpg icon on the toolbar.

    The icon is selected, and the cursor changes to image26100557.jpg.

  2. Move the cursor towards the comment to be deleted. When the cursor changes to image265.jpg, click the mouse.

    The comment is deleted.

  3. When finished, click image25900558.jpg again.

Deleting Peaks

You can delete or restore a peak that has been auto-integrated.

A deleted peak cannot be used for a calibration point, FQ blank, and so on. Also, the peak is not displayed in a report.

To delete a peak, complete the following steps:

  1. Click DeletingPeaks.png on the toolbar.

    "Peak deletion" mode is enabled.

  2. Move the cursor to the peak that has been auto-integrated.

    The cursor changes to DeletingPeaksCursor.png.

  3. Click the peak.

    Auto-integration for the peak is cleared and a red cross is displayed above the peak.

    The cells corresponding to the peak in the batch list become blank.

Follow the steps below to restore a peak that was deleted using this function.

  1. Click DeletingPeaks00559.png on the toolbar.

    "Peak deletion" mode is enabled.

  2. Move the cursor to the deleted peak.

    The cursor changes to RestoringPeakCursor.png.

  3. Click the peak.

    The peak is restored.

Manual integration

You can manually integrate the peaks that are not set as analytes in the data analysis method displayed in the upper Chromatogram pane (Extracted Ion Chromatograms). You can also change and manually integrate the peaks that have been set as the analytes in the data analysis method in the lower pane.

 

Peaks that were processed by autointegration are shown in green lines. Unknown peaks that cannot be processed by autointegration are shown in white.

 

To perform manual integration, follow these steps:

  1. If image36700560.jpg on the toolbar is on the toolbar is selected (when a toolbar icon is selected, it is highlighted in a darker shade), click to deselect (inactivate).

    If image36500561.jpg is not selected, click to select (activate).

    The Chromatogram is displayed in the upper pane, while an expanded view of the selected peak is displayed in the lower pane.

  2. To manually integrate in the lower pane, use PreviousElement.png and NextElement.png to select the peak to be integrated manually.
  3. Click the image345.jpg icon on the toolbar. To perform manual integration along the peak, also click image371.jpg.

    Move the cursor towards the peak in the lower row. The cursor changes to a image347.jpg.

  4. Click and drag the start or end black circle to specify the integration range.

    manual_integration_new.png 

    Manual Integration

  5. Release the mouse button.

    Integration is performed and the Batch Table is updated for the analyte that was integrated manually in the lower pane.

  6.  

    If you integrate unknown peaks, you will not be able to view the integration results in the Batch Table. Right-click on the Chromatogram pane and select [Export Area Percent Report] from the context menu to view the integration results. For peaks that have been manually integrated, "(M)" is displayed in the peak #.

     

  7. When finished, click image34500564.jpg again.
  8. To clear the manual integration results for a known peak, click the image351.jpg icon on the toolbar.

 

Move the cursor near the manually integrated peak and click it when the cursor changes to CHclearsekibun.png. Note that you cannot clear all manually-integrated peaks together at once. Clear them one by one.

 

Displaying multiple chromatograms

This function is not available on the Chromatogram pane that is displayed in the [Method Editor] window. Use this function on the Chromatogram pane displayed in Batch-at-a-Glance. By default, the Chromatogram pane (Option) displays the chromatogram for one data only. However, you can also overlay chromatograms for multiple data for comparison.

Displaying multiple chromatograms

To display multiple chromatograms, follow these steps:

  1. Click the image353.jpg icon on the toolbar.

    The [Change Chart] dialog box is displayed.

    change_chart_new.png 

    [Change Chart] Dialog Box

    The following settings are available:

    • For the current sample, show the chromatogram for the current mass (default)

      Select [Current Masses of Current Sample].

    • For the current sample, show the chromatogram for multiple masses

      Select [Multi Masses of Current Sample], and then in the [View] column, mark the check boxes for the masses to display. To display the peaks upside down, mark the check boxes in the [Invert] column.

    • For the current mass, show the chromatogram for multiple samples

      Select [Multi Samples of Current Mass], select the data in the dialog box that is displayed, and click <Add>. In the [View] column, mark the check box for the sample data to display. To display the peaks upside down, mark the check boxes in the [Invert] column.

  2. When finished, click <OK>.

    The chromatograms are overlaid.

    CH_Overlay.png 

    Superimposed Chromatogram

    To display respective chromatograms in separate graphs, click CH_SeparateIcon.png the toolbar.

    CH_SeparateView.png 

    Superimposed Chromatogram

Resetting the display of multiple chromatograms

To reset the overlaid chromatograms to the original state, follow these steps:

  1. Click the image35300565.jpg icon on the toolbar.

    The [Change Chart] dialog box is displayed.

  2. Select [Current Mass of Current Sample] and click <OK>.

    The original view is restored.

Displaying the S/N ratio

You can test the reliability of the chromatography results by calculating the signal-to-noise (S/N) ratio of the peaks.

To calculate and display the S/N ratio of the peaks, you must first set the noise range on the Chromatogram pane. Setting the range on the upper row (Extracted Ion Chromatogram) of the Chromatogram pane applies the specified noise range to all compounds. Setting the range on the lower row (Selected Peak) applies the specified range to the particular compound only.

 

You may only specify one noise range for each chromatogram. To change the noise range, simply specify a different range.

 

Displaying the S/N ratio on the labels

To display the S/N ratios on the peaks, configure the peak labels to display S/N ratios.

To display the S/N ratios on the labels, do the following:

  1. Click PeakLabel.png on the toolbar.

    The [Chromatogram Information] dialog box is displayed.

  2. If the [S/N ratio] check box on the [Chromatogram Information] dialog box is not marked, mark it.

    chromato_label_SN.png 

    Chromatogram Information Dialog Box

  3. Click [OK].

    The S/N ratios will now be displayed in the peak labels.

    Proceed to “Setting the noise range”.

Setting the noise range

On the Chromatogram pane, set the noise range to use to calculate the S/N ratios.

  1. Click noise.png on the toolbar.

    The icon is selected, and the cursor changes to NoiseCur.png.

  2. Drag the mouse to specify the desired range.

    The noise range for calculating the S/N ratios is now set.

    NoiseDrag.png 

    Setting the Noise Range

  3. Click [Process Batch] from the [Batch Option] group on the [Home] tab.

    S/N ratios are displayed in the peaks in the lower row.

    SNRatioDisplay.png 

    S/N Ratio Display

 

 

Clearing the noise range

To clear the specified noise range, do the following:

  1. If noise00566.png on the toolbar is not selected, click it.

    The icon is selected, and the cursor changes to NoiseCur00567.png.

  2. Move to the noise range to clear.

    The cursor changes to NoiseDelCur.png.

    SN_Del.png 

    Clearing the Noise Range

  3. Click the mouse.

    The noise range is cleared, as are the S/N ratios displayed in the peaks in the lower row.

  4. Click [Process Batch] from the [Batch Option] group on the [Home] tab.

    The analysis results are updated.

Exporting the Signal to Noise report

This function is not available on the Chromatogram pane that is displayed in the [Method Editor] window. Use this function on the Chromatogram pane that is displayed in Batch-at-a-Glance. The Chromatogram Signal to Noise Report can be created as a file.

The following file formats are available:

However, you can select an Excel file only when the Excel is installed.

The created report can be edited and graphed in the associated application.

 

When using ECM, OpenLab Server Products, Workstation Plus, or SDA, the displayed dialog box and the save file destination differ from standard MassHunter operations. For more information, refer to "Operations When Database Systems Are Used" in "Reference".

 

  1. Set the noise range as described in “Setting the noise range”.
  2.  

    Be sure to process the batch after you set the noise range. 

     

  3. Right-click on the Chromatogram pane (Option).

    The context menu is displayed.

  4. Select [Export Signal to Noise Report].

    The [Signal to Noise Report] dialog box is displayed.

    SNReport.png 

    [Signal to Noise Report] dialog box

  5. Enter a file name, then click <Save>.

    The Signal to Noise Report opens in the associated application.

    signal_to_noise_excel_report.png 

    Signal to Noise Report (when using Excel)

  6. After you view the report, close the application.

Updating methods

If the current peak values do not match the values in the analysis method, for example due to the deterioration of the column, the analysis method can be updated with the values for the current peak. This function lets you make changes to the analysis method from the Chromatogram pane, so that you do not have to open the [Method Editor] window.

 

This function is useful for qualifier ions.

 

  1. Use PreviousElement.png and NextElement.png, to select the analysis method to update.
  2. Click Update.png.

    The [Update Method for Current Peak] dialog box is displayed.

    UpdateDialog.png 

    The values for the analysis method and the current peak are displayed.

  3. Mark the Update check box for the items to update.
  4. Click [OK].

    The selected items are updated on the analysis method.

  5. Click [Process Batch] from the [Batch Option] group on the [Home] tab.

    The analysis results are updated.

Exporting the Area Percent report

The Chromatogram Area Percent Report can be created as a file.

The following file formats are available:

However, you can select an Excel file only when the Excel is installed.

The created report can be edited and graphed in the associated application.

 

By selecting the submenu in the context menu, you can create the following three types of Area Percent reports.

 

 

When using ECM, OpenLab Server Products, Workstation Plus, or SDA, the displayed dialog box and the save file destination differ from standard MassHunter operations. For more information, refer to "Operations When Database Systems Are Used" in "Reference".

 

  1. From the Batch table, select one or more samples or masses used to create a report.

    The contents to be selected vary depending on the type of the Area Percent report you want to create. For details, refer to the above NOTE.

  2. Right-click on the Chromatogram pane (Option).

    The context menu is displayed.

  3. Select [Export Area Percent Report].
  4. According to the type of the report you want to create, select [All], [All Quantitated Masses], or [Selected Mass] form the submenu.

    The [Area Percent Report] dialog box is displayed.

    CHAreaPercent.png 

    [Area Percent Report] dialog box

  5. Enter a file name, then click <Save>.

    The Area Percent Report opens in the associated application.

    CHAreaPercentExcel.png 

    Area Percent Report (when using Excel)

  6. After you view the report, close the application.

Tabulating raw and CPS data

A CSV format table can be created based on the raw or CPS data for the chromatogram.

A listing of the mass count or CPS for each time unit are created and displayed in the associated application.

The created table can be edited and graphed in the associated application.

  1. Right-click on the Chromatogram pane (Option).

    The context menu is displayed.

  2. Select [Tabulate Chart], and select one of the following commands from the submenu.
    • [Raw Data to CSV]: Exports the raw data in CSV format, and then opens it.
    • [CPS Data to CSV]: Exports the CPS data in CSV format, and then opens it.
  3. The associated application starts, and the raw or CPS data for the chromatogram is displayed.

    chromatorawexcel.gif 

    Raw Data Table for the Chromatogram (when using Excel)

  4. Close the application.

    The [ICP-MSICP-QQQ Data Analysis] window is displayed.

Copying chromatograms

Chromatograms can be copied to the Clipboard. They can then be pasted into documents created with other applications. For more information, refer to “Copying the graphs” under “Common Graph Operations”.

Generating the Quick Batch Report

To print a summary of the analysis results, print a Quick Batch Report. For more information, refer to “Generating the Quick Batch Report” under “Common Data Analysis Operations”.

Printing chromatograms

Chromatograms can be printed. For more information, refer to “Printing the panes” under “Common Pane Operations”.

Exporting chromatograms

Chromatograms can be exported in various graphics file formats. For more information, refer to “Exporting the graphs” under “Common Graph Operations”.