Set the acquired elements to the batch.
When a tuning mode is selected for which [MS/MS] is specified for [Scan Type], the following tabs are displayed.
Precursor Ion Scan Tab
You can set the Q2 mass for performing a Precursor Ion Scan. By double-clicking the tab, the [Set Q2] dialog box is displayed. By entering a Q2 mass in the text box and then clicking the [OK] button, the settings of a Precursor Ion Scan for all masses are added in the mass table.
Product Ion Scan Tab
You can set the Q1 mass for performing a Product Ion Scan. By double-clicking the tab, the [Set Q1] dialog box is displayed. By entering a Q1 mass in the text box and then clicking the [OK] button, the settings of a Product Ion Scan for all masses are added in the mass table.
Set Mass Shift Tab
You can set a Mass Shift. By selecting a mass row for which a Mass Shift is set on the mass table and then clicking this tab, the [Set Mass Shift] table is displayed.
Tune Mode
Select a tuning mode.
Mass table
When you click on the mass scale to select a mass, the selected mass is displayed in the right table.
[Set Mass Shift] table
Predefined Shift
By selecting a shift amount from Predefined Shift, the shift amount is added to each mass and the shifted masses are set in the Q2 column.
NH3 Cluster
When the shift amount for the NH3 cluster is selected, the shift amount is added to each mass and the shifted masses are set in the Q2 column.
Custom Shift
Custom shift of masses and/or Mass ranges separated by commas. For example, type "1,18,28-30".
[Clear All]
Clears all target elements that are set in the tuning mode.
[Go to Periodic Table]
When you click <Go to Periodic Table>, the display style is switched to a periodic table.
[OK]
Applies the settings.
[Cancel]
Cancels the settings.