Data Analysis for FullQuant Analysis

Quantitative analysis is a method for determining the concentration of the elements in an unknown sample.

A calibration curve is first created based on a standard sample that has a known concentration of elements. The concentration of the elements in the unknown sample is then calculated by comparing with this reference data.

For more information, see Reference > Example of Analysis Samples.

With MassHunter Workstation, data that are acquired by adding ISTD (Internal STanDard) elements or by the standard addition method can also be used for more accurate analysis results.

To perform quantitative analysis using data acquired with MassHunter Workstation for the unknown sample and the standard sample, perform analysis using a Data Analysis Method for quantitative analysis.

This section describes how to create a Data Analysis Method for quantitative analysis, perform the analysis, and check the analysis results.

To perform quantitative analysis with the optional Chromatographic Application, refer to “Data Analysis for Chromatogram (Option)”.

Loading data into the batch folder

Open a Batch Folder, and load the acquired data to be analyzed by completing the following steps.

You may also load the data after you create the Data Analysis Method.

  1. In the [ICP-MSICP-QQQ Data Analysis] window, open a Batch Folder.

    For more information, refer to “Opening the Data Analysis window”, “Creating a batch folder”, and “Loading a batch result” under “Common Data Analysis Operations”.

  2. In the [ICP-MSICP-QQQ Data Analysis] window, load the appropriate data files, including the standard samples for the calibration curves, the background data, and the unknown sample data.

    For more information, refer to “Loading the data” under “Common Data Analysis Operations”.

    The data files are displayed in the Batch Table pane.

    BatchTable_Sample.png 

    Batch Table Pane with Samples

Sample type setup

Specify the Sample types and calibration curve level.

This screen shot shows examples of sample types and calibration curve levels.

BatchTable_SampleType.png 

Batch Table Pane with sample types specified

  1. Check that the correct Sample types are displayed in the Type column of the Batch Table pane. If they are incorrect, change the setting.
    • Background sample (for background subtraction): Bkgnd
    • FQ blank sample (for blank concentration subtraction): FQBlk
    • Blank sample for calibration curves: CalBlk (set Level column to “1”)
    • Standard sample for calibration curves: CalStd (set Level column to “2”, “3”, ... in ascending order of concentration)
    • Unknown sample: Sample

Creating a Data Analysis method for fullquant

To create a Data Analysis Method for FullQuant, follow these steps:

  1. Click [Edit] from the [Method] group on the [Home] tab.

    The [Method Editor] window is displayed.

    MethodEditorWindow.png 

    [Method Editor] Window

  2. In the Method Development Tasks pane, click [Data Analysis Method] from the [Set up Basic Information] group.

    The Data Analysis Method pane is displayed.

    DAmethodPane.png 

    Data Analysis Method Pane

  3. Configure the settings in the Data Analysis Method pane as follows:
    FQDAmethodTable.png 

    Data Analysis Method Table

    • Mark the check box for [FullQuant Analysis].
    • The [Analysis Mode] list displays the analysis mode for the acquired data. The analysis mode can be changed.

      Spectrum: Spectrum analysis mode (includes Multi Tune)

      Timechart: Time resolved analysis mode or Time Program mode

      Chromatogram: For details on the Chromatogram mode, refer to “Data Analysis for Chromatogram”.

     

    “Spectrum” can be selected for data that was acquired in spectrum analysis mode. “Timechart” or “Chromatogram (Option)” can be selected for data that was acquired in Time Resolved analysis mode.

    If not analyzed yet, all of these options can be selected. (Chromatogram is optional.)

     

    • Select the background subtraction method from the [Bkg Subtraction if Exists] list. This option is not displayed for Chromatogram mode.

      When [Bkg Subtraction if Exists] is checked, select “Bkgnd” for the sample type of the background sample.

    • To perform interference correction:

      Select the correction method from the [Interference Correction] list. This option is not displayed for Chromatogram mode.

    For more information about these settings, refer to “Data Analysis Method pane”.

  4. In the Sample/Batch Template table, configure the template for the analysis results report.

    There are two types of templates:

    • Sample Template: Report of sample data
    • Batch Template: Report of the entire batch

    Click the image110.jpg on the right end of the box to select a template, as necessary.

    FQtemplateTable.png 

    Sample/Batch Template Table

  5.  

     

  6. On the Method Development Tasks pane, click [FullQuant] from the [Set up Analysis Parameters] group.

    The Calibration pane is displayed.

    CalibrationPane_MS.png

    CalibrationPane_QQQ.png

    Calibration Pane

  7. If the analytes are not listed in the Calibration pane, try one of the following:

  8. In the Calibration pane, configure each analyte as follows:
    • Click the image11000094.jpg at the right end of the Tune Mode column, and select a Tune Mode.
    • Click the image11000095.jpg at the right end of the Analyte/ISTD column, and assign the element to be an analyte or an ISTD element.
  9. For more information, refer to “Calibration pane”.

  10. In the Calibration pane, set the calibration curve parameters.

    For more information about these settings, refer to “Calibration pane”.

  11. Configure the following settings as necessary:
  12. On the Method Development Tasks pane, click [Return to Batch-at-a-Glance] from the [Validate/Return] group.

    If an error is found in the Data Analysis Method, the error contents are displayed in the Method Error List pane. Correct the error.

    If no errors are found, then a confirmation dialog box is displayed.

    noerror_MS.png

    Confirmation Dialog Box

  13. Click <OK>.
  14. On the Method Development Tasks pane, click [Return to Batch-at-a-Glance] from the [Validate/Return] group.
  15. When asked whether to update the Data Analysis Method, click <Yes>.

    Proceed to the analysis process.

Executing analysis

If a new Data Analysis Method was created or the settings were changed, do an analysis. To run an analysis, complete the following step:

  1. Click [Process Batch] from the [Batch Option] group on the [Home] tab.

    Analysis is performed using the settings in the Data Analysis Method, and the analysis results are displayed in the Batch Table pane.

    Proceed to Checking/Correcting the analysis results.

    FQDAWindow_MS.png

    FQDAWindow_QQQ.png

    Analysis Results

Checking/Correcting the analysis results

This section describes the procedures that are used to check and correct the FullQuant analysis results.

Checking the Batch Table pane

The Batch Table pane displays the concentration and count for each element in the samples. Various tools are provided that let you easily review the analysis results.

For more information on viewing and using the Batch Table pane, refer to “Batch Table Pane Operation”.

Checking the spectrum pane

If the Analysis Mode is “Spectrum”, the Spectrum pane is displayed.

For more information on viewing and using the Spectrum pane, refer to “Spectrum Pane Operation”.

Checking the timechart pane

If the Analysis Mode is “Timechart”, the Timechart pane is displayed.

For more information on viewing and using the Timechart pane, refer to “Timechart Pane Operation”.

Checking the calibration curve pane

The Calibration Curve pane displays the calibration curves for each element in the samples.

For more information on viewing and using the Calibration Curve pane, refer to “Calibration Curve Pane Operation”.

Checking the ISTD stability graph pane

When you use internal standard correction, the recovery (%) of each ISTD element is displayed in the ISTD Stability Graph pane. If the recovery value is <100%, then the analysis results may be less reliable.

For more information on viewing and operating the ISTD Stability Graph pane, refer to “ISTD Stability Graph Pane Operation”.

Saving the analysis results

For more information, refer to “Saving the analysis results” under “Common Data Analysis Operations”.

Generating the Quick Batch Report

To print a summary of the analysis results, print a Quick Batch Report. For more information, refer to “Generating the Quick Batch Report” under “Common Data Analysis Operations”.

Generating the analysis results report

For more information, refer to “Generating the analysis results report” under “Common Data Analysis Operations”.

Closing the Data Analysis window

For more information, refer to “Closing the Data Analysis window” under “Common Data Analysis Operations”.